Header menu link for other important links
X
Microstructure and wear properties of LENS® deposited medical grade CoCrMo
, C.K. Esplin, B.E. Stucker
Published in
2008
PMID: 17990083
Volume: 19
   
Issue: 5
Pages: 2105 - 2111
Abstract
There is a growing interest in metal-on-metal bearing surfaces for orthopedic implants. Although some success has been achieved in applications like hip implants which involve a large contact area, non-conforming joints, such as knees, have proved more difficult. The current work examines the applicability of a novel additive manufacturing process, Laser Engineered Net Shaping (LENS®, the registered trademark and service mark of Sandia National Laboratories and Sandia Corporation), for producing CoCrMo implants. A series of experiments were conducted to determine the optimum parameters for deposition of CoCrMo. Microstructural studies, hardness tests, and dry sand/rubber wheel abrasive wear tests were conducted on the LENS® deposits. The results showed that metallurgically sound deposits can be produced using the LENS® process under optimized conditions. The hardness of the LENS® deposited CoCrMo was found to be comparable to that of standard CoCrMo wrought material; however, wear tests indicated that LENS® deposits were considerably less resistant to abrasive wear than wrought CoCrMo. The reasons for this behaviour are discussed based on microstructural observations. © 2007 Springer Science+Business Media, LLC.
About the journal
JournalJournal of Materials Science: Materials in Medicine
ISSN09574530