Measurement of tendon velocities using vector tissue Doppler imaging: A feasibility study
, L. Bellini, L. Prosser, C. Stanley, D. Bland, K. Alter, D. Damiano, S. Sikdar
Published in
2010
PMID: 21096066
Pages: 5310 - 5313
Abstract
We have developed a vector Doppler ultrasound imaging method to directly quantify the magnitude and direction of muscle and tendon velocities during movement. The goal of this study was to evaluate the feasibility of using vector Tissue Doppler Imaging (vTDI) for estimating the tibialis anterior tendon velocities during dorsiflexion in children with cerebral palsy who have foot drop. Our preliminary results from this study show that tendon velocities estimated using vTDI have a strong linear correlation with the joint angular velocity estimated using a conventional 3D motion capture system. We observed a peak tendon velocity of 5.66±1.45 cm/s during dorsiflexion and a peak velocity of 8.83±2.13 cm/s during the passive relaxation phase of movement. We also obtained repeatable results from the same subject 3 weeks apart. Direct measurements of muscle and tendon velocities may be used as clinical outcome measures and for studying efficiency of movement control. © 2010 IEEE.
About the journal
Journal2010 Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBC'10